Contact Us

Got Questions? We're Here to Help! Contact Us.

2173 Salk Avenue, Suite 250 (#2046)
Carlsbad, CA, 92008
United States

+1 (510) 899-1340

Welcome to Service AI, your trusted source for accessories, spare parts, and certified pre-owned laboratory instrumentation tailored for scientists and startups. We provide reliable, current equipment at competitive prices, along with essential support for instruments no longer made or backed by their original manufacturers.

Our expertise spans leading brands like Beckman Coulter and Agilent, and we are committed to high-quality standards through certified maintenance, repair, and professional installation services. We believe our installation services provide the confidence and assurance you need when purchasing instrumentation from a reliable and trustworthy source.

At Service AI, we don't conduct the science—we empower scientists with reliable laboratory instrumentation and spare parts that are guaranteed to perform. We trust in scientists' expertise to determine the instrumentation they need to advance their research.

With more than 25 years of experience, our dedicated team prides itself on delivering only the highest quality products. Every instrument we offer has been thoroughly tested and certified by manufacturer-trained engineers. We ensure the functionality of our products with a Full Performance Guarantee and provide proof of performance documentation when applicable.

At Service AI, we also recognize the importance of meticulous shipping and safe delivery. Our products are handled with the utmost care to ensure they arrive in perfect condition. For those who require it, we offer professional installation services to help integrate our equipment seamlessly into your operations.

Purchase our reliable scientific laboratory equipment, and we’ll provide seamless installation service worldwide for your startup or research needs.

Advance your scientific endeavors with confidence: rely on us for recertified, trustworthy instruments at reasonable prices. It’s a smart choice made simple!

GC-MSD System.jpg

Chromatography Guide

In-depth guides, applications, and technical insights for mastering leading chromatography systems, including Agilent Gas Chromatographs and Mass Spectrometers.

Filtering by Tag: Baseline Noise

Guide to: Gas Chromatography (GC) Troubleshooting

Service AI

In capillary gas chromatography, the instrument and the column function as a single, integrated system rather than independent components. A deficiency in any part of this system typically manifests as a chromatographic difficulty, requiring a logical and systematic approach to identify the source. This Lab Pundit guide provides a technical overview of column anatomy, performance variables, and diagnostic procedures for maintaining system integrity.

Anatomy of a Capillary Column

A capillary column is a complex device comprising three distinct layers, each critical to performance:

  • Fused Silica Tubing: Synthetic quartz containing less than 1 ppm metallic impurities, serving as the structural foundation.

  • Polyimide Coating: An outer waterproof barrier that fills flaws in the tubing to provide strength and durability.

  • Stationary Phase: A polymer (typically silicone or polyethylene glycol) coated on the inner wall that governs the separation process.

Critical Performance Variables

Maintaining optimal chromatography requires monitoring technical thresholds that define the "health" of the system:

  • Column Capacity: The maximum sample amount injectable before significant peak distortion occurs. Overloading is typically indicated by a "sharkfin" shape with a leading edge.

  • Column Bleed: The background signal resulting from the continuous degradation of the stationary phase. While inherent to all columns, excessive bleed can be accelerated by oxygen exposure at high temperatures.

  • Temperature Limits: Columns feature an isothermal limit for continuous use and a higher program limit for short durations (typically <10 minutes).

Diagnostic Categories for Troubleshooting

Performance issues generally fall into eight technical categories. Systematic identification is the most efficient route to a solution.

1. Baseline Disturbances

Baseline irregularities often point to detector or gas supply issues.

  • Spiking: Usually caused by particulate matter in the detector or loose electrical connections.

  • Noise: Can result from contaminated injectors, air leaks in ECD/TCD systems, or incorrect combustion gas flow rates.

  • Drift: Upward drift often indicates system contamination or damaged stationary phase. Downward drift is common during the conditioning of a new column.

2. Irregular Peak Shapes

Peak morphology provides direct insight into the chemical environment of the column.

  • Tailing: Often a symptom of an active or contaminated injector liner, poorly cut column ends, or a polarity mismatch between the solute and stationary phase.

  • Split Peaks: Can be caused by erratic injection technique, poor column installation in the injector, or coelution of multiple compounds.

  • Reduced Size: Frequently points to a leak in the injector, a partially plugged syringe, or a split ratio that is set too high.

Technical Best Practices for Maintenance

Consistent maintenance prevents the most common causes of system failure, such as contamination and phase damage.

  • Column Cutting: Use a carbide or diamond-tipped pencil to scribe the polyimide, ensuring a clean, square cut to avoid adsorptive losses.

  • Gas Purification: High-purity carrier gases are essential. Use moisture and oxygen traps in series to protect against irreversible oxidative damage to the stationary phase.

  • Conditioning: Always verify carrier gas flow and the absence of leaks before heating a column to its conditioning temperature to prevent permanent phase damage.

  • Solvent Rinsing: For bonded and cross-linked phases, solvent rinsing can remove semi-volatile residues and restore performance.